Hold it any way up. Point it at the target face-on. The Helios reads the same whether it is level, sideways, tilted or upside down. What matters is the angle between the device and the hot surface, not the angle between the device and the floor. For spot sizes, the laser dots and the reticle, see Optics and sighting in Using the Device.
Not needed
Holding the device level, or square to the ground.
What counts
Looking at the hot surface face-on, within about 45°.
Orientation
The reading does not depend on how the device is held: level, sideways, upside down, up or down. The two laser dots turn with the device, and the measured spot is always exactly halfway between them.
Level
On its side
Upside down
Pointing up
Pointing down
Viewing angle
Angle is measured from face-on. Looking straight at the surface is 0°. Skimming along it is 90°. Stay within about 45° and you are fine on almost any surface. Beyond 60°, move yourself if you can.
Shiny surfaces may read low.
The reading drops, the spot stretches, and reflections of hot surroundings grow.
The zones depend on the surface. Rough and oxidised surfaces (scale, oxide, refractory, most castings) tolerate steeper angles. Polished, glassy and liquid-metal surfaces need to be viewed closer to face-on.
Spot size
The round measuring spot lands on the surface as an oval, longer in the direction you are tilted and the same width. Its length grows as 1 / cos(angle). Drawn to scale below for the 20 mm spot at 1–2 m, on a narrow target 30 mm wide.
Face-onround
20 mm
45°1.4× longer
28 × 20 mm
60°2× longer
40 × 20 mm
80°5.8× longer
115 × 20 mm
Tips
The target must be larger than the spot: about 20 mm at 1–2 m, 30 mm at 3 m, 60 mm at 5 m, 140 mm at 10 m. At 1–3 m, if the target fills the gap between the two green dots, it fills the spot.
The spot is exactly halfway between the two green dots. On screen it sits on the centre line, a little below the middle (the camera is just above the measuring path), and nearer the middle for farther targets. The reticle shows that path; see Where the device is measuring.
Smoke, steam, dust, flames and dirty sight-glasses make the reading low. Move to a clearer view or wait for a gap. Through a furnace opening, look straight through so its edge does not cut into the spot.
Dirt or a film on the measuring window reads low. Clean it with a soft lens cloth, and take the protective cover off before measuring.
Default 1.00, settable 0.01–1.20. Use the value for the material and its surface state; the Emissivity page lists typical values. A wrong setting is a fixed offset on every reading.
Average, held steady, for a flickering or noisy target. Maximum, sweeping slowly, to catch the hottest, cleanest area of a scaled or patchy surface. Minimum to find the coolest point.
Between 1 and 3 m the spot is smallest (20–30 mm). Farther away is fine for large targets, as long as the target still fills the spot.
The Helios works in surroundings of 10–60 °C. Do not stand in the hot air rising from a furnace opening or ladle; measure from the side or further back, still near face-on to the surface.