Operation

Holding and Aiming

Hold it any way up. Point it at the target face-on. The Helios reads the same whether it is level, sideways, tilted or upside down. What matters is the angle between the device and the hot surface, not the angle between the device and the floor. For spot sizes, the laser dots and the reticle, see Optics and sighting in Using the Device.

Not needed

Holding the device level, or square to the ground.

What counts

Looking at the hot surface face-on, within about 45°.

Any way up

The reading does not depend on how the device is held: level, sideways, upside down, up or down. The two laser dots turn with the device, and the measured spot is always exactly halfway between them.

Helios held level, aimed face-on at a hot target. Reading is correct.

Level

Helios rolled onto its side, aimed face-on at a hot target. The laser dots turn with it. Reading is correct.

On its side

Helios held upside down, grip on top, aimed face-on at a hot target. Reading is correct.

Upside down

Helios tilted up about 45 degrees, aimed face-on at an overhead hot part. Reading is correct.

Pointing up

Helios pointing steeply down into a ladle of molten metal, looking nearly face-on at the melt surface. Reading is correct.

Pointing down

Figure 1 — Any way up. The reading does not depend on how the device is held. The small squares show the face of the target as you see it: the measured spot (dashed) sits halfway between the two green dots, whichever way the dots turn.

The one thing to watch: angle to the surface

Angle is measured from face-on. Looking straight at the surface is 0°. Skimming along it is 90°. Stay within about 45° and you are fine on almost any surface. Beyond 60°, move yourself if you can.

Viewing angle zones above a hot surface A fan of angles measured from the face-on line, which points straight out of the surface. Green from 0 to 45 degrees: good on any surface. Amber from 45 to 60 degrees: fine on rough or oxidised surfaces, shiny surfaces may read low. Red from 60 to 90 degrees: avoid. A Helios at 30 degrees gets a tick; one at 75 degrees gets a cross. 0° face-on 45° 60° 90°
  • 0–45°: good on any surface
  • 45–60°: OK on rough or oxidised surfaces

    Shiny surfaces may read low.

  • Beyond 60°: avoid

    The reading drops, the spot stretches, and reflections of hot surroundings grow.

The zones depend on the surface. Rough and oxidised surfaces (scale, oxide, refractory, most castings) tolerate steeper angles. Polished, glassy and liquid-metal surfaces need to be viewed closer to face-on.

Illustration: smooth hot glass at 1000 °C. An example worked from standard optics, not a Helios specification. The error stays small until about 60°, then grows fast. Reading low by:
0°
none
45°
under 2 °C
60°
about 6 °C
70°
about 16 °C
80°
about 50 °C
Why the Helios is forgiving. It measures at a short wavelength, 1 µm, where an emissivity error becomes a much smaller temperature error than on long-wavelength instruments. As a rule of thumb, at 1000 °C a 5 % emissivity error is about 6 °C.
Figure 2 — Angles are measured from face-on, not from the surface: 0° is the good direction. Rules of thumb from standard optics. How much a surface is affected depends on the material, its roughness and any oxide layer.

At an angle, the spot stretches

The round measuring spot lands on the surface as an oval, longer in the direction you are tilted and the same width. Its length grows as 1 / cos(angle). Drawn to scale below for the 20 mm spot at 1–2 m, on a narrow target 30 mm wide.

Face-on: spot is a 20 millimetre circle, inside the 30 millimetre target.

Face-onround
20 mm

45 degrees: spot is an oval 28 by 20 millimetres, still inside the target.

45°1.4× longer
28 × 20 mm

60 degrees: spot is an oval 40 by 20 millimetres and spills off the 30 millimetre target, so it reads low.

60°2× longer
40 × 20 mm

80 degrees: spot is an oval 115 by 20 millimetres, far longer than the target is wide.

80°5.8× longer
115 × 20 mm

Narrow hot target, 30 mm wide Spot on the target Spot spilling off: averages in cooler background, reads low
Figure 3 — At an angle the round spot becomes an oval: 2× longer at 60°. On a wire, rod, thin stream or narrow strip, a stretched spot spills off the edge and the reading drops.

Tips for a good reading

Fill the spot

The target must be larger than the spot: about 20 mm at 1–2 m, 30 mm at 3 m, 60 mm at 5 m, 140 mm at 10 m. At 1–3 m, if the target fills the gap between the two green dots, it fills the spot.

Aim with the dots and the screen

The spot is exactly halfway between the two green dots. On screen it sits on the centre line, a little below the middle (the camera is just above the measuring path), and nearer the middle for farther targets. The reticle shows that path; see Where the device is measuring.

Keep a clear line of sight

Smoke, steam, dust, flames and dirty sight-glasses make the reading low. Move to a clearer view or wait for a gap. Through a furnace opening, look straight through so its edge does not cut into the spot.

Clean window, cover off

Dirt or a film on the measuring window reads low. Clean it with a soft lens cloth, and take the protective cover off before measuring.

Set the emissivity for the material

Default 1.00, settable 0.01–1.20. Use the value for the material and its surface state; the Emissivity page lists typical values. A wrong setting is a fixed offset on every reading.

Match the mode to the job

Average, held steady, for a flickering or noisy target. Maximum, sweeping slowly, to catch the hottest, cleanest area of a scaled or patchy surface. Minimum to find the coolest point.

Pick a comfortable distance

Between 1 and 3 m the spot is smallest (20–30 mm). Farther away is fine for large targets, as long as the target still fills the spot.

Look after the device too

The Helios works in surroundings of 10–60 °C. Do not stand in the hot air rising from a furnace opening or ladle; measure from the side or further back, still near face-on to the surface.